Kollef Marin H et al., 2008-08-20: Multicentre randomised trial in 2003 ventilated patients (the NASCENT trial) finding a significant reduction in microbiologically confirmed ventilator-associated pneumonia and delayed onset with a silver-coated endotracheal tube.
Patients receiving a silver-coated endotracheal tube had a statistically significant reduction in the incidence of VAP and delayed time to VAP occurrence compared with those receiving a similar, uncoated tube.
Abstract
Kollef Marin H et al., 2008-08-20: Multicentre randomised trial in 2003 ventilated patients (the NASCENT trial) finding a significant reduction in microbiologically confirmed ventilator-associated pneumonia and delayed onset with a silver-coated endotracheal tube. C6 Study type: prospective, randomized, single-blind, controlled multicentre trial; 54 centres, 9417 screened, 2003 randomized. Country: North America (USA). Silver form: silver-coated endotracheal tube (product not named in the abstract text captured; the FDA clearance for the Agento tube is recorded separately). Coating concentration not stated in the abstract. Comparator: otherwise identical uncoated high-volume low-pressure tube. Outcomes: VAP incidence 4.8% versus 7.5% among patients intubated 24 h or longer (P=.04), relative risk reduction 35.9%; no significant differences in mortality, ICU or hospital stay. Funder: not stated in the abstract text captured; clinicaltrials.gov NCT00148642. Retraction check: none. Study type: randomized trial.