Magnetic Resonance Imaging-Guided Breast Biopsy in Korea: A 10-Year Follow-Up Experience (2021)

Peer-Reviewed Study · Cha SY, Ko EY, Han BK, Ko ES, Choi JS, Park KW, Lee JE · July 8, 2021

Cha SY et al., 2021-07-08: Ten-year single-centre retrospective review of MRI-guided vacuum-assisted breast biopsy in 74 scheduled procedures (65 performed): the false-negative rate was 4.3% (2/46 benign diagnoses later found malignant), DCIS was underestimated in 43% (3/7) and ADH in 50% (1/2); the authors set their 4.3% against 17.4%-23.2% false-negative rates in earlier reports.

The false-negative rate in our study was very low, compared with the false-negative rate of 17.4%-23.2% in previous reports

Discussion

Cha SY et al., 2021-07-08: Ten-year single-centre retrospective review of MRI-guided vacuum-assisted breast biopsy in 74 scheduled procedures (65 performed): the false-negative rate was 4.3% (2/46 benign diagnoses later found malignant), DCIS was underestimated in 43% (3/7) and ADH in 50% (1/2); the authors set their 4.3% against 17.4%-23.2% false-negative rates in earlier reports. Retrospective cohort; n=74 lesions/patients scheduled, 65 biopsies performed, 38 of 48 benign lesions followed for a median 5.8 years; South Korea (Samsung Medical Center, Sungkyunkwan University, Seoul); funding: no funding statement in the article; declared conflicts: 'The authors declare that they have no competing interests'; device: 9-gauge vacuum-assisted biopsy (make/model not stated), typically six samples; comparator: subsequent surgical excision (n=27) and MRI/mammography follow-up. Full text read through PMC. The paper's own conclusion also states that close follow-up MRI may not be necessary after a concordant benign result, which is recorded here as the authors' position. Retraction status: PubMed record type 'Journal Article'; no CommentsCorrections entry and no retraction or expression of concern in the record. Study type: uncontrolled series.