CT-guided transthoracic needle aspiration biopsy of pulmonary nodules: needle size and pneumothorax rate (2003)

Peer-Reviewed Study · Geraghty PR, Kee ST, McFarlane G, Razavi MK, Sze DY, Dake MD · November 2003

Geraghty PR et al., 2003-11: Retrospective review of 846 consecutive CT-guided transthoracic needle aspiration biopsies using an 18- or 19-gauge outer coaxial stabilizing needle through which a smaller aspiration needle or automated biopsy gun was passed. Pneumothorax occurred in 226/846 patients; 124/324 (38%) with 18-gauge needles versus 121/522 (23%) with 19-gauge needles (P<.001). Coaxial needle size and patient age significantly affected pneumothorax rate; diagnostic accuracy was 96% with 18-gauge and 92% with 19-gauge needles.

Pneumothorax occurred in 124 (38%) of 324 patients who underwent procedures with 18-gauge needles and in 121 (23%) of 522 patients who underwent procedures with 19-gauge needles (P <.001).

Geraghty PR et al., 2003-11: Retrospective review of 846 consecutive CT-guided transthoracic needle aspiration biopsies using an 18- or 19-gauge outer coaxial stabilizing needle through which a smaller aspiration needle or automated biopsy gun was passed. Pneumothorax occurred in 226/846 patients; 124/324 (38%) with 18-gauge needles versus 121/522 (23%) with 19-gauge needles (P<.001). Coaxial needle size and patient age significantly affected pneumothorax rate; diagnostic accuracy was 96% with 18-gauge and 92% with 19-gauge needles. Item: Geraghty et al. Study type: retrospective cohort; 846 consecutive procedures; United States (Stanford University Medical Center). Funding: not stated in the retrieved abstract. Conflicts of interest: not stated in the retrieved abstract. Device: 18- or 19-gauge outer coaxial stabilizing needle with a smaller aspiration needle or automated biopsy gun (make/model not given in the abstract). Comparator: 18-gauge versus 19-gauge coaxial needle size, with univariate and multivariate regression on age, sex, smoking, lesion size, core gun use and number of passes. Retraction check: Europe PMC record for PMID 14595149 lists no retraction; one 'comment in' item only. Access note: abstract-only record retrieved. Study type: controlled study.