Effect on oximetry of dyes used for sentinel lymph node biopsy. (2004)

Peer-Reviewed Study · Piñero A, Illana J, García-Palenciano C, Cañizarese F, Canteras M, Cañadillas V, Durán E, Parilla P · November 2004

Piñero A et al., 2004-11: Prospective randomised study in 32 women having sentinel node biopsy for breast cancer plus 6 controls: both methylene blue and isosulfan blue disturbed pulse oximetry, but only isosulfan blue reached statistical significance, and the desaturation was factitious in both cases.

The 2 dyes interfered with the peripheral saturation reading, but only isosulfan blue showed significant differences.

Piñero A et al., 2004-11: Prospective randomised study in 32 women having sentinel node biopsy for breast cancer plus 6 controls: both methylene blue and isosulfan blue disturbed pulse oximetry, but only isosulfan blue reached statistical significance, and the desaturation was factitious in both cases. C8. Study type: prospective randomised comparative trial with a non-randomised control group. N: 32 patients (16 MB, 16 isosulfan blue) + 6 controls. Country: Spain (Hospital Universitario Virgen de la Arrixaca, Murcia, per author group). Dye dose/route: sentinel node injection, peritumoural; doses not in the abstract. Funder/conflicts: none declared in the record. Comparator: isosulfan blue and no-dye controls; outcomes pulse oximetry vs blood-gas co-oximetry at 15, 30, 60 min. Partly because the finding is that MB interferes less than the comparator dye, and the setting is breast surgery, not prostate surgery. Study type: randomized trial.