Interference of methylene blue with CO-Oximetry of hemoglobin derivatives. (1997)

Peer-Reviewed Study · Gourlain H, Buneaux F, Borron SW, Gouget B, Levillain P · June 1997

Gourlain H et al., 1997-06: Laboratory study of six CO-Oximeters showing that methylene blue absorbs in the same 550-700 nm region as haemoglobin derivatives and perturbs total haemoglobin and derivative readings, so these instruments should not be used on blood containing MB. The laboratory basis for false methaemoglobin readings during MB administration.

We conclude, therefore, that these six CO-Oximeters should not be used to determine the concentration of hemoglobin derivatives in blood samples containing MB.

Gourlain H et al., 1997-06: Laboratory study of six CO-Oximeters showing that methylene blue absorbs in the same 550-700 nm region as haemoglobin derivatives and perturbs total haemoglobin and derivative readings, so these instruments should not be used on blood containing MB. The laboratory basis for false methaemoglobin readings during MB administration. C8. Study type: laboratory methods study (Clin Chem 1997;43(6 Pt 1):1078-80, letter-length article). N: not applicable (spiked blood aliquots at 10, 25, 50, 100 mg/L MB). Countries: France (Hopital Fernand Widal, Paris) with a US co-author. Methylene blue: stock 10 g/L solution (Pharmacie Centrale des Hopitaux de Paris) diluted into blood to 10-100 mg/L, chosen to match plasma levels after 5-25 mL of a 10 g/L intravenous dose. Funder/conflicts: none declared in the text. Comparator: NaCl 9 g/L control aliquots and six different CO-Oximeters. Quote from the full text opened via the DOI; the PubMed record has no abstract. The article's figure labels and the greater-than glyphs extract unreliably, so the quoted sentence was chosen to avoid them.