Determining susceptibility of routine clinical biochemistry assays to methylene blue interference. (2025)

Peer-Reviewed Study · Stevic I, Kllapi T, Bhayana V, Jasani A, Rutledge AC · May 2, 2025

Stevic I et al., 2025-05-02: Laboratory study spiking patient specimens and donor blood with methylene blue across twelve Roche cobas chemistry assays, co-oximetry and urinalysis; MB caused measurable interference at low mg/L concentrations and made co-oximetry incalculable at 50-70 mg/L, while dipstick urinalysis was affected from 5 mg/L.

However, a urine MB concentration of ≥5 mg/L interfered with most urinalysis parameters on the cobas u601 analyzer. Automated comments based on colour detection were introduced to make clinicians aware of potential MB interference with urinalysis.

Stevic I et al., 2025-05-02: Laboratory study spiking patient specimens and donor blood with methylene blue across twelve Roche cobas chemistry assays, co-oximetry and urinalysis; MB caused measurable interference at low mg/L concentrations and made co-oximetry incalculable at 50-70 mg/L, while dipstick urinalysis was affected from 5 mg/L. C8. Study type: laboratory interference study (in vitro spiking of leftover patient specimens and healthy-donor blood). N: not applicable. Country: Canada (authors Stevic, Khlapi, Bhayana, Jasani, Rutledge – Alberta laboratory medicine group). Methylene blue: 10 or 100 mg/L initial spikes, then dose-response; venous blood spiked for co-oximetry. Funder/conflicts: 'The authors declare that they have no known competing financial interests'. Instruments: Roche cobas c502/c702, cobas u601, Werfen GEM Premier 4000. Comparator: water-spiked controls. Note: the assay panel is general chemistry and urinalysis; the abstract does not list total or free PSA among the twelve assays, so this paper does not by itself document PSA assay interference. Study type: lab.