Development and validation of an infrared-artificial intelligence software for breast cancer detection (2023)

Peer-Reviewed Study · Martín-Del-Campo-Mena E, Sánchez-Méndez PA, Ruvalcaba-Limon E, Lazcano-Ramírez FM, Hernández-Santiago A, Juárez-Aburto JA, Larios-Cruz KY, Hernández-Gómez LE, Merino-González JA, González-Mejía Y · April 27, 2023

Martín-Del-Campo-Mena E et al., 2023-04-27: This developer study trained AI classifiers on a proprietary infrared-image database and validated the best one in 349 patients against mammography under a double-blind design, reporting 94.87% sensitivity, 72.26% specificity and 99.12% negative predictive value for the thermography software, against 100% sensitivity and 97.10% specificity for mammography in the same patients; it proposes the software as a complementary screening tool.

The infrared-AI software here developed shows high BC sensitivity (94.87%) and high NPV (99.12%).

Abstract, Conclusions

Martín-Del-Campo-Mena E et al., 2023-04-27: This developer study trained AI classifiers on a proprietary infrared-image database and validated the best one in 349 patients against mammography under a double-blind design, reporting 94.87% sensitivity, 72.26% specificity and 99.12% negative predictive value for the thermography software, against 100% sensitivity and 97.10% specificity for mammography in the same patients; it proposes the software as a complementary screening tool. C8 Martín-Del-Campo-Mena 2023, Explor Target Antitumor Ther 4:294-306 (PubMed epub 2023-04-27). Diagnostic accuracy study: AI development on a proprietary database (20,022 images from 3,337 patients) then prospective-style double-blind clinic validation in 349 patients, 39 with confirmed breast cancer; country: Mexico (FUCAM A.C., a specialised breast pathology centre); funding: not stated in the article's funding field, but the authors declare a relationship with Hearthcore SAPI de CV (PASM, AHS, JAJA, LEHG, JAMG and YGM as associates; EMDCM as medical advisor with non-financial support), which the article labels a potential competing interest; device: FLIR T650sc infrared camera with proprietary infrared-AI software; comparator: mammography read in the same patients, with mammography, ultrasound and biopsy as reference standard. The software's own numbers are below its mammography comparator on both sensitivity (94.87% vs 100%) and specificity (72.26% vs 97.10%), and the authors propose complementary use rather than substitution, so the row is filed PARTLY SUPPORTS. PubMed record shows no retraction notice (checked 2026-11-22). Study type: controlled study.