NADP Analytical Laboratory et al., 2026-02-23: Analytical-laboratory QA plan for the samples: supply blanks for buckets, bags, bottles and filters, lot acceptance criteria against method detection limits, and the blank/spike/duplicate schedule, i.e. the laboratory half of the QA/QC chain.
The lot testing criteria states, for each NTN analyte, that the mean of at least 10 samples per lot must be < NTN MDLN and none of the supply blanks in the batch tested may exceed 3 times the NTN MDLN.
p. 28
NADP Analytical Laboratory et al., 2026-02-23: Analytical-laboratory QA plan for the samples: supply blanks for buckets, bags, bottles and filters, lot acceptance criteria against method detection limits, and the blank/spike/duplicate schedule, i.e. the laboratory half of the QA/QC chain. text_file=nadp_labqa2026.txt Effective 2/23/2026, 87 pages. Section 8.2.3 'NTN/AMoN Lot Testing Criteria'. Section 8.1 lists supplies routinely checked for contamination, including NTN bucket lids, 1 L HDPE bottles and sampling bags. Study type: regulatory record.